Skip to main content
Fig. 7 | Journal of Cardiovascular Magnetic Resonance

Fig. 7

From: A deep learning pipeline for automatic analysis of multi-scan cardiovascular magnetic resonance

Fig. 7

Bland–Altman plots of late gadolinium enhancement (LGE) scar percentage between automated and manual measurements (a), as well as between two measurements by a same human observer (b). The middle line denotes the mean difference (bias) and the two dashed lines denote ± 1.96 standard deviations from the mean. The plot a shows a mean difference of 1.1% with 95% limits of agreement being from − 14.8 to 13.5%. The plot b shows a mean difference of 4.4% with 95% limits of agreement being from − 10.6 to 19.4%

Back to article page