Table 1 The CoV, accuracy, and precision of T1 mapping and T2 mapping of T1MES phantom

From: Reference ranges of myocardial T1 and T2 mapping in healthy Chinese adults: a multicenter 3T cardiovascular magnetic resonance study

Vial no.

T1 mapping

T2 mapping

CoV (%)

Accuracy (ms)

Precision (ms)

CoV (%)

Accuracy (ms)

Precision (ms)

RR interval = 600 ms

   

Vial A

0.9

15

0.9

12.1

5

11.0

Vial B

3.6

134

3.2

6.3

14

4.7

Vial C

0.5

6

0.5

2.3

74

1.5

Vial D

0.8

25

0.7

4.8

6

4.2

Vial E

5.0

215

4.3

6.6

14

4.9

Vial F

7.9

169

7.2

17.2

183

6.9

Vial G

1.0

10

0.9

2.8

2

2.7

Vial H

2.7

82

2.5

6.1

10

5.0

Vial I

19.8

5

20.1

12.2

46

9.2

RR interval = 900 ms

    

Vial A

0.5

17

0.4

4.0

6

3.5

Vial B

3.6

119

3.3

4.3

13

3.4

Vial C

0.8

6

0.8

2.2

72

1.4

Vial D

0.6

27

0.6

4.1

6

3.7

Vial E

4.4

189

3.8

5.2

12

4.0

Vial F

7.1

122

6.7

10.0

159

4.8

Vial G

0.7

11

0.7

2.3

2

2.2

Vial H

2.2

79

2.1

3.9

9

3.3

Vial I

19.9

4

20.1

8.9

51

6.5

  1. The coefficient of variation (CoV) was defined as the ratio of the SD to the mean. Accuracy was defined as average of absolute difference between T1 (T2) measurements obtained from T1 (T2) SE and T1 (T2) mapping. Precision was defined as the average of the SD of T1 (T2) measurements obtained with T1 (T2) mapping